Should all Coagulation Specimens be PPP?

A reprise of a post from a few years ago:

At the Mayo Medical Laboratories Coagulation Testing Quality Conference in Minneapolis, Dr. Dorothy Adcock Funk of Esoterix reported on the 2008 revision of Clinical and Laboratory Standards Institute (CLSI) Standard H21-A5, Collection, Transport, and Processing of Blood Specimens for Testing Plasma-Based Coagulation Assays and Molecular Hemostasis Assays. One interesting comment she made is that it is not necessary to prepare platelet-poor plasma (PPP; plasma with a platelet count less than 10,000/uL) for routine coagulation testing on fresh plasma. She reconfirmed, however, that plasmas for freezing and lupus anticoagulant (LA) testing must be PPP. Bioactive platelet granule materials are released upon freezing, including platelet factor 4 (PF4), which neutralizes heparin. In LA testing, platelets provide high concentrations of membrane phospholipid, which may neutralize low-titer LAs.

At UAB we centrifuge all our coagulation specimens, routine and special, at 2000g for 15 minutes, producing PPP in a single spin. Dr. Larry Brace at Edwards Hospital, Naperville, IL, double-spins all specimens using the Stat-spin Express centrifuge. Our reasoning is that even though it isn’t necessary to prepare PPP for routine tests, our laboratory staff often get calls for special add-ons using existing specimens. Further, we may choose to reflex unexpectedly prolonged partial thromboplastin time (PTT) assays to our LA profile.
I’d like to know what you do about PPP preparation; all specimens, or just those for freezing and LA profiles? Please answer by providing your comments below. Geo.

Comments (1)
Blog
JoeLamb
May 7, 2009 10:06am

We just perform routine coag. We spin with a Statspin using
We just perform routine coag. We spin with a Statspin using a sigle cycle. Special test are referred out and those usually require frozen plasma. The chemistry lab double spins those render PPP, because they handle all the referred test. If special test are added on, chemistry will spin those too, and freeze. We periodically check the Statspin performance by performing a platelet count on spun plasma samples.

Leave a Reply